This isolated output 4-20mA industrial transmitter, a sturdy microprocessor-based instrument, measures conductivity and temperature. It transmits conductivity measurements via a two-wire power supply loop. The industrial transmitter has many user friendly features—all of which are easily accessible via the water resistant membrane keypad. Ideal for connecting to a PLC for advanced control systems, the transmitter can also be connected to a recorder for a lowcost documentation system. The instrument housing is designed for panel or wall mounting.
- Six selectable conductivity ranges for a wide range of measurements
- Isolated Output prevents output interference caused by electrical noise or leakage from the input source
- Out-of-range Output Current Indicator indicates an out-of-range measurement by emitting a 3.8mA output current
- Hold Output Current Indicator to indicate ‘HOLD’ mode during calibration or set up mode.
- Scaleable 4-20 mA Current Output
- Primary and secondary display for simultaneous display of Conductivity and Temperature values
- Easy Push Button Operation
- Automatic (with Pt 100) or manual Temperature Compensation
- Panel or Wall Mount, Nema 4x, IP65 Weatherproof Housing
Specifications:
1) Conductivity Ranges:
R1: 0-19.99 μS/cm, 0.01 μS/cm resolution
R2: 0-199.9 μS/cm, 0.1 μS/cm resolution
R3: 0-1999 μS/cm, 1μS/cm resolution
R4: 0-10.00 mS/cm, 0.01 mS/cm resolution
R5: 0-19.99 mS/cm, 0.01 mS/cm resolution
R6: 0-199.9 mS/cm, 0.1 mS/cm resolution
2) Cell Constants
k=0.1, R1-R5
k=1.0, R1-R6
k= 10.0, R2-R6
3) Accuracy
±1% of full scale reading in respective range
4) Temperature
0 °C to 100 °C (32 °F to 212 °F)
5) Temperature Resolution
0.1 °C (0.1 °F)
6) Relative Temperature Accuracy
±0.5 °C (±0.9 °F)
7) Sensor Pt100
(3 wire / 2 wire)
8) Temperature Compensation
Auto/Manual
9) Output
4.0-20.0mA (Galvanically isolated)
10) Cell Constant User-Selectable
(0.10 to 10.00)
11) Temperature Coefficient
0 to 10% (selectable)
12) Normalization Temperature
20.0 °C to 25 °C
13) Display
Dual Display Custom LCD
14) Power Requirements
12 to 24V DC, maximum 600 Ohm load
15) Dimension
96 x 96 x 66 mm (3.78 x 3.78 x 2.6 inch)
16) Weight
210 g
17) Certifications
CE
Documentation: